Two probes on a 2.0 kΩ source
Drawn above at its default parameters, which is almost never how an essay calls it. A placement states the numbers that essay is arguing about, so the figure a reader meets is about that argument rather than about the generator — 98% of the placements on this site pass one, and the phase that raised that number from 12% found eight captions describing a figure the page was not showing.
At those defaults the edge it states is 1% wrong at 6.79 kHz with a 1× probe — the right-hand slot
of the caption strip, which on this site is never used for anything else, and which is read
back out of the drawing above rather than out of the code that wrote it.
It belongs to Measurement, which is a circuit on a circuit, which is to say a change to it is a
change to lib/figures/instruments.js.
It takes a slider on source resistance (ohms) with 6 settings,
and every one of them has passed the same assertions as the frame above — a figure whose
circuit stops doing what its caption says at any setting stops the build.
Called by 5 essays
which is the blast radius of changing it
The probe is part of the circuit
A one-to-one oscilloscope probe on a two-kilohm source gives a reading that is one per cent wrong at 6.8 kHz. Not because the instrument is inaccurate — it is reading correctly — but because the hundred and fifteen picofarads at the end of the cable are across the node, and above that frequency the trace on the screen is a picture of a circuit that only exists while the probe is attached.
A divider with two ratios
Put capacitance in a resistive divider and it divides by resistance at direct current and by capacitance at high frequency, and those are two different numbers unless one equation holds. The adjustable trimmer on every oscilloscope probe exists for that single equation, and the square wave on the instrument's front panel is a display of which of the two ratios is currently winning.
The instrument's own rise time
Rise times add in quadrature, so ten per cent of inflation needs an instrument 2.18 times faster than the edge. That constant contains no instrument. Measured on the solved network it is 2.79 for a one-pole front end, 3.97 for two, 4.87 for three and 5.62 for four — the rule is optimistic at every pole count, which is the wrong direction for a rule of thumb to err in.
The rejection the parts have
Two essays measured the architecture: four resistors decide an instrumentation amplifier's rejection, two do not, and the answer is the one-amplifier figure plus twenty log of the first stage's gain — exactly, with amplifiers of infinite rejection. Give each amplifier its own and something unobvious happens: two matched but individually mediocre parts cost nothing at all, because their error is a common-mode signal at the difference stage and is rejected there. What costs is the difference between them, and it sets a ceiling with no gain in it.
The probe that takes a tenth
A ten-to-one probe buys an order of bandwidth for a tenth of the signal, and on a two-kilohm source the bandwidth is exact: 6.79 kHz becomes 69.2 kHz. The tenth of the signal is not a tenth of the signal-to-noise ratio. Solved resistor by resistor, the noise referred to the tip goes from 1.782 µV to 55.78 µV — a factor of 31.3 — because the divider that does the attenuating is nine megohms and a megohm, and √(n(n−1)kT/C) on the cable's own capacitance has no source resistance in it at all.