Bridge — the series
-
The bridge that is linear near one point
The expression for a Wheatstone bridge's output is V delta over four, and the solve says it is V delta over two times two plus delta — low by half the fractional change, exactly, at every change tested. So the tangent is worth one per cent only up to a two per cent change and a tenth of a per cent only up to two tenths. Driving the bridge from a current source instead halves the departure at every point and doubles both edges, which is a change of one component and no change at all to the four resistors being measured.
-
The leads that are in the bridge
A strain gauge on the end of two long wires cannot be told from a strain gauge under load: both leads in the changing arm is bit for bit the same netlist as a quarter bridge whose fractional change is larger by twice the lead over the gauge, which at 350 milliohms on a 350 ohm gauge is 999 microstrain that is not there. Moving one of those leads into the arm beside it leaves the output at exactly zero for every lead resistance drawn, and turns a twenty-kelvin drift of 76.8 microstrain into 0.077. That factor is two over the strain being read — 999 at a thousand microstrain and 9990 at two hundred — and it costs 0.100 per cent of sensitivity.
-
The two leads nobody counts
The leads that are in the bridge moved a lead out of the changing arm and turned a 999-microstrain error into nothing. The two leads carrying the excitation are still there, and they scale every reading: 0.35 ohms each on a 350-ohm bridge is 1,995 parts per million, exactly −2r/(R + 2r), the same on a full bridge as on a quarter one, and drifting 0.153 microstrain over twenty kelvin — twice what the three-wire fix left behind. Two more wires brought back from the bridge's own terminals leave 0.00035 parts per million. So does exciting the bridge with a current, which needs no extra wires at all.
-
The carrier a cable allows
A carrier-frequency amplifier removes an instrument's offset drift by exciting the bridge with alternating current, and the cable that was a resistance at direct current becomes a capacitance. Both output terminals see the same arms and the same capacitance, so the lag cancels in the difference: ten metres matched leaves 0.15 degrees at a hundred kilohertz where each terminal has 6.3. What does not cancel is the mismatch between the two conductors, arriving multiplied by the ratio of the excitation to the signal — two thousand at a thousand microstrain — so one per cent of it takes the allowed carrier from 205 kilohertz to 1.8.
-
The capacitance that is already a difference
A screened pair has capacitance from each conductor to the screen, which cancels in a bridge's difference when the two match, and capacitance between the conductors, which is a difference from the start and cannot cancel. Ten metres of it between a 350 Ω quarter bridge and its carrier amplifier allows 18.2 kHz for a 200 ppm loss, against 205 kHz for the screen alone. The two phases are ωR·ΔC/δ and ωR·C, the same slope with one divided by the signal, so a screen mismatch of δ·Cp/Cs on the gauge's own terminal cancels the pair's lag exactly: 0.10% at 1000 µε takes the limit to 143 kHz. That is what a carrier amplifier's capacitance balance finds when it nulls the quadrature — and because the mismatch's phase changes sign with the strain and the pair's does not, the balance holds in tension above half the strain it was set at and never in compression.
-
What a calibration at the carrier removes
The phase a cable puts on a carrier-excited bridge is a loss of 1 − cos φ, and a loss that is the same at every strain is a gain error — which a shunt calibration at the operating carrier exists to remove. Calibrated there, a synchronous reading of a quarter bridge through ten metres of matched screened pair allows 121 kHz for 200 ppm over ±2000 µε, where the phase measure allowed 16.5; the capacitance between the conductors, which no matching could improve, stops mattering. Where the reference is taken — at the instrument or from sense wires at the bridge — moves that by 1.2 per cent. What calibration cannot touch is a screen mismatch: its leak is a fixed voltage, not a fixed gain, and a hundredth of a per cent of it takes 121 kHz back to 11.5.
-
The sensor that is the capacitance
Read a capacitive displacement sensor through the same kind of cable that limited a strain-gauge bridge's carrier, and the roles exchange. On the bridge, a 1% mismatch in the cable's capacitance was a quadrature leak that grew as the square of the carrier and could be escaped by a lower one. On a sensor of two 10 pF plates it is 3.0% of full-scale displacement, in phase, at every carrier from 100 Hz to 1 MHz — because a capacitance between an excitation conductor and the sense conductor is in parallel with the measurand and is a sensor capacitance. No detector phase separates it and no carrier escapes it; only geometry does. Screen each conductor separately and there is nothing left to mismatch.
-
The reference that should be late
A carrier amplifier's reference is taken from the excitation because that is what the output is meant to be in phase with, and nothing says that is where it belongs. A screen mismatch's leak arrives late by the common-mode node's lag and the difference's together, 525 ns on ten metres of pair and a 350 Ω bridge, so a reference on the excitation reads sin 3.7° of it at 20 kHz: 0.607 per cent of error at a tenth of a per cent of mismatch, and 72.6 parts per million with the reference 3.7° late. The null is a delay rather than a phase and does not move with the mismatch, and a delayed reference allows about five times the carrier on any mismatched screen. How exactly it must be held narrows as mismatch times carrier.