Bridge — the series
-
The bridge that is linear near one point
The expression for a Wheatstone bridge's output is V delta over four, and the solve says it is V delta over two times two plus delta — low by half the fractional change, exactly, at every change tested. So the tangent is worth one per cent only up to a two per cent change and a tenth of a per cent only up to two tenths. Driving the bridge from a current source instead halves the departure at every point and doubles both edges, which is a change of one component and no change at all to the four resistors being measured.
-
The leads that are in the bridge
A strain gauge on the end of two long wires cannot be told from a strain gauge under load: both leads in the changing arm is bit for bit the same netlist as a quarter bridge whose fractional change is larger by twice the lead over the gauge, which at 350 milliohms on a 350 ohm gauge is 999 microstrain that is not there. Moving one of those leads into the arm beside it leaves the output at exactly zero for every lead resistance drawn, and turns a twenty-kelvin drift of 76.8 microstrain into 0.077. That factor is two over the strain being read — 999 at a thousand microstrain and 9990 at two hundred — and it costs 0.100 per cent of sensitivity.
-
The two leads nobody counts
The leads that are in the bridge moved a lead out of the changing arm and turned a 999-microstrain error into nothing. The two leads carrying the excitation are still there, and they scale every reading: 0.35 ohms each on a 350-ohm bridge is 1,995 parts per million, exactly −2r/(R + 2r), the same on a full bridge as on a quarter one, and drifting 0.153 microstrain over twenty kelvin — twice what the three-wire fix left behind. Two more wires brought back from the bridge's own terminals leave 0.00035 parts per million. So does exciting the bridge with a current, which needs no extra wires at all.