Current sensing — the series
-
The ammeter that is a resistor
Every direct measurement of a current is a measurement of a voltage across something the current was made to flow through, so the instrument has two errors pointing opposite ways: a larger shunt changes the current, a smaller one leaves less for the amplifier's offset to be compared with. Written in the burden voltage they are the burden over the supply and the amplifier's offset over the burden, and the best of them is at the geometric mean — 7.75 millivolts on a twelve-volt rail, with a worst-case error of 0.129 per cent. Neither number contains a resistance, and neither contains the current: ten milliamps and a hundred amps want the same 7.75 millivolts.
-
The ammeter that is not in the circuit
A shunt measures a current by putting a resistance in the circuit, and every objection to it follows from that. A current transformer puts nothing in the circuit at all — a thousand-turn secondary reflects twelve microhms into the primary — and charges for it in a different currency: no response at direct current, a ratio error that stops falling at one minus the coupling, and a phase error of half a degree at fifty hertz that costs eighteen per cent of a power reading at a power factor of 0.05.
-
The optimum that hands back a bandwidth
The best burden voltage across a shunt is 7.75 mV and contains neither the current nor the resistance, which is what made it worth having. A shunt has two nanohenries whatever it is made of, so holding the burden fixed fixes the resistance at u*/I — and the corner R/2πL then falls in exact proportion to the current: 6.16 MHz at a tenth of an ampere, 616 kHz at one, 6.16 kHz at a hundred. A resistor and a 258.2 pF capacitor across it, found by search on the solved response, flatten the reading to 8×10⁻⁵ per cent across six decades.
-
The rail that is an input error
The best burden voltage across a shunt is the geometric mean of the amplifier's offset and the supply, and moving the shunt to the high side does not change that form — it changes what the offset is. A hundred-decibel amplifier on a twelve-volt rail turns the rail into 120 µV of equivalent input error, twenty-four times its own five, so the optimum moves from 7.75 mV and 0.129 per cent to 38.73 mV and 0.644. The two are equal only at 128 dB, and with the common-mode term removed the optimum returns to the low-side value exactly.
-
The resistance that depends on the reading
Three of a shunt's errors are free of the current being measured, which is the whole content of the burden-voltage optimum. The fourth is not: the shunt dissipates, warms, and its resistance rises — so the divisor the reading uses is a function of the reading. Solved as a fixed point it agrees with R₀/(1 − αθI²R₀) to 2×10⁻¹⁶, and along the optimum, where the dissipation is I·u* rather than I²R, the error is the FIRST power of the current: 7.75 ppm at an ampere, 775 at a hundred, fitted exponent 1.0007.