Series

Four-terminal — the series

4 essays on one idea, from the one that introduces it to the one that assumes the rest.
  1. Measuring with 50 mΩ of lead in each wire. computed by solving, not by drawing at 61 resistances, twice each. The two-wire arrangement measures the leads too, so its error is 2×50 mΩ over whatever is being measured: one per cent at 10 Ω, and 10000% at 1 mΩ. The four-wire arrangement senses on a separate pair that carries almost no current, and its error stays under 1.0e-2% across the whole range.

    Two terminals measure the leads as well

    Fifty milliohms in each lead makes a two-wire measurement one per cent high at ten ohms, ten per cent high at one ohm, and a hundred per cent high at a tenth. Not approximately — the reading is the resistance plus the leads, and below about ten ohms most of what is being reported is the wire between the instrument and the thing.

    part 1 · instruments
  2. A coupled pair as a two-port, at k = 0.8. computed by solving, not by drawing. Each port driven in turn with the other open, four solves, and the four impedance parameters read out. The diagonal terms measure each winding's own inductance — 10.0000 mH and 40.0000 mH against 10 and 40 — and both transfer terms measure the mutual inductance, 16.0000 mH against k√(L₁L₂) = 16.0000. The two transfer terms agree to 2.83e-16, which is reciprocity — a property of the device rather than of the measurement, and the first thing a coupling stamped into the wrong row would break. Neither of this site's two standing checks can see it: a coupling adds no current and dissipates nothing.

    Two ports from two one-ports

    An inductor is a one-port: one impedance, one number. Two of them coupled is a two-port, and the four impedance parameters that describe it are recovered here by four solves — each port driven with the other open, the definition read literally. Two of the four come out equal to 2.8 × 10⁻¹⁶, which is reciprocity, and is the first property a coupling stamped into one row instead of two would break.

    part 2 · magnetics
  3. Four wires against a 10 MΩ voltmeter. computed by solving, not by drawing at 81 resistances, twice each, with a voltmeter of 10 MΩ and 50 mΩ in every lead. The four-wire error is not zero: it is the voltmeter's own divider, −(R + 2R_lead)/(R + 2R_lead + R_m), which grows with the resistance being measured rather than shrinking. The two-wire error is that same quantity plus the leads, so it passes through zero at 1000 Ω — where the reading is right to 1.8e-12 while the four-wire reading is 0.0100% low — and above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.

    The voltmeter four wires do not remove

    A four-terminal measurement is described everywhere as removing the leads from the answer. It moves them. What is left is the voltmeter's own input resistance, and it grows with the resistance being measured rather than shrinking: with a ten-megohm voltmeter and fifty milliohms of lead, the four-wire reading is 0.0100 per cent low at a kilohm, where the two-wire reading is exactly right — 1.8 × 10⁻¹² — because its lead error and its loading error cancel. Above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.

    part 3 · instruments
  4. A milliohm read four-terminal with a drifting thermoelectric EMF: the error against the test current, read three ways, and the shunt's own heating. computed by solving, not by drawing: an EMF of 0.5 µV settled plus 1 µV relaxing with a minute's time constant, read thirty seconds after a disturbance with readings a second apart; the shunt 20 ppm/K and 20 K/W. At 10 mA one polarity is out by 11.1%, a reversal by 501 ppm, the three-reading delta by 4.14 ppm. Every EMF error falls as 1/I; the heating, 0.400 ppm at an ampere, rises as I². The best current is 11 A for one polarity (149 ppm in all), 1.8 A reversed (4.08 ppm) and 0.37 A with the delta (0.167 ppm).

    The voltage four wires carry in

    Four wires remove a resistance measurement's leads by keeping the sense loop free of current, and that loop is exactly where a thermoelectric voltage lives. Ten milliamps through a milliohm is ten microvolts, so every microvolt where a copper lead meets a brass terminal moves a single-polarity reading by a tenth. Reversing the current removes any EMF that holds still and leaves the change over one interval; the + − + delta leaves the second difference; four readings leave the third — 501 ppm, 4.14 and less at 10 mA while a disturbed junction settles. Against the shunt's own heating each scheme has a best current, and the least error rises only as the cube root of the shunt's temperature coefficient: the reading method moves the error by orders of magnitude, the alloy by a cube root.

    part 4 · instruments

All series