A one-to-one oscilloscope probe on a two-kilohm source gives a reading that is one per cent wrong at 6.8 kHz. Not because the instrument is inaccurate — it is reading correctly — but because the hundred and fifteen picofarads at the end of the cable are across the node, and above that frequency the trace on the screen is a picture of a circuit that only exists while the probe is attached.
The probe that takes a tenth
A ten-to-one probe buys an order of bandwidth for a tenth of the signal, and on a two-kilohm source the bandwidth is exact: 6.79 kHz becomes 69.2 kHz. The tenth of the signal is not a tenth of the signal-to-noise ratio. Solved resistor by resistor, the noise referred to the tip goes from 1.782 µV to 55.78 µV — a factor of 31.3 — because the divider that does the attenuating is nine megohms and a megohm, and √(n(n−1)kT/C) on the cable's own capacitance has no source resistance in it at all.