Concept

Measurement error — where it appears

The difference between what an instrument reads and what the circuit would do without it. Computed here by solving the network twice, with the instrument present and absent, because it is a property of the pair rather than of either one on its own.

Named by 19 essays across 8 fields — each of them below, with the objects they name alongside it.

A clean board is 14.0 mV of error, and a guard makes it 1.0 nV. computed by solving, not by drawing. Two boards differing by one wire, solved with the leakage present and again with it removed, so the number plotted is the leakage's own contribution and nothing else. Unguarded, a teraohm across the laminate from a 15 V rail through a 1 GΩ source is 13.99 millivolts — and a humid morning takes that resistance down two decades, which is the left-hand end of this axis. Guarded, the ring is held at the input's own potential by the amplifier, so what is across the leakage is the amplifier's own error and the result is 1.00 nanovolts. The guarded line is flat in the rail and proportional to the signal: the offset has become a gain error of 1.00 parts per billion, and the reading is low by it rather than high: the ring sits a little below the input, so the last of the leakage pulls the input down.

The current that does not reach the input

A teraohm across a board from a fifteen-volt rail is fourteen millivolts of error through a gigohm source, and a humid morning takes that resistance down two decades. A ring held at the input's own potential leaves a nanovolt — proportional to the signal rather than to the rail, so an offset has become a gain error of one part in a billion — and the same wire multiplies the input resistance by the loop gain, which makes it 10¹⁸ Ω at direct current and 10¹² Ω at a megahertz.

instruments · Guarding
A tenth of a per cent of lead is 999 µε of strain that is not there. computed by solving, not by drawing. A 350 Ω quarter bridge at zero load, against the resistance in each lead. With both leads in the changing arm the output is 4.99500 mV at 350 mΩ — an apparent fractional change of 0.1998%, which at a gauge factor of 2 is 999 microstrain. With one lead in that arm and one in the arm beside it the output is zero to the last bit, at every lead resistance drawn. The lower panel is what the second arrangement costs: the sensitivity falls as 1/(1 + Rlead/R), which is 0.100% at the same lead and is a calibration constant rather than a drift.

The leads that are in the bridge

A strain gauge on the end of two long wires cannot be told from a strain gauge under load: both leads in the changing arm is bit for bit the same netlist as a quarter bridge whose fractional change is larger by twice the lead over the gauge, which at 350 milliohms on a 350 ohm gauge is 999 microstrain that is not there. Moving one of those leads into the arm beside it leaves the output at exactly zero for every lead resistance drawn, and turns a twenty-kelvin drift of 76.8 microstrain into 0.077. That factor is two over the strain being read — 999 at a thousand microstrain and 9990 at two hundred — and it costs 0.100 per cent of sensitivity.

applied · Bridge
A whole number of steps: 1, 2, 3 and 4 agree to 7% and a step and a half is 1.9× worse. computed by solving, not by drawing. The upper panel is the share of the quantisation error sitting in harmonics of the input against the amount of dither added — the upper curve the largest share anywhere on the amplitude sweep, with the spread across the five tones each point averages drawn as a bar, and the lower curve that sweep's mean. Undithered the worst is 76.5 per cent. It falls steeply up to one whole step (4.70 per cent at three quarters, 0.34 at one) and then stops improving — but only AT whole steps. The sweep means at 1, 2, 3, 4 steps are 0.283, 0.297, 0.297, 0.301 per cent, flat to 7 per cent; at 1.5, 2.5, 3.5 they are 0.526, 0.345, 0.311, each above both whole steps beside it. The lower panel is what each costs in signal-to-noise ratio, with 10·log₁₀(1 + L²) drawn through it — the measurement is that curve to 0.118 dB everywhere, so the price is known in advance and only the benefit has to be measured. The choice is a corner and a comb: nothing here is minimised, something stops improving, and between the places where it has stopped it is worse again.

The dither that is a decision

One whole least significant bit is quoted everywhere as the dither, which makes a decision look like a constant. Swept, the axis is a corner and a comb. An eighth of a step leaves 64.5 per cent of the error locked to the signal and one whole step leaves 0.34; above that the sweep mean is 0.283, 0.297, 0.297 and 0.301 per cent at one, two, three and four steps and 0.526 at a step and a half, which fails at exactly the small amplitudes dither exists for. The price is 10·log₁₀(1 + L²) to 0.118 of a decibel, and four steps cost 12.41 for nothing.

digital · Quantisation
The "offset" a switch leaves moves by 1.863 mV across a volt of signal. computed by solving, not by drawing. The charge a switch leaves on a 1 pF hold capacitor, as the input it was sampling changes. It is W·L·C_ox·(V_gs − V_th) with the clock's high level for the gate, so it falls as the input rises for two reasons at once: less gate drive, and a threshold that rises with the source potential through the body effect. The dashed line is the part that does not know what the signal is — the clock coupling through the gate overlap capacitance, -1.649 mV — and it is the only part of this that is honestly an offset. The curve stops where the switch does: an n-channel device passes nothing within a threshold of its own gate, and the model refuses rather than returning a zero.

The offset that knows the signal

The charge a switch leaves on the capacitor it was sampling is the gate oxide capacitance times the channel area times the overdrive, and both terms in that overdrive depend on the input — one directly and one through the body effect. So the 5.67 mV a data sheet would call an offset moves by 1.86 mV across a volt of signal, which is a gain error of 0.186 per cent and only 3.4 µV of anything else. Opening the summing-node switch first divides the gain error by the amplifier's own gain, exactly, and what is left goes from twenty-nine times the sampled-noise floor to nineteen times below it.

filters · Switched-capacitor
Two leads nobody counts cost 1995 parts per million, and two more leads remove all of it. computed by solving, not by drawing. The error in the reported strain against the resistance in each of the two excitation leads, at 1000 µε on a quarter bridge of 350 Ω excited at 10 V. With four wires the instrument takes the excitation to be the supply's voltage, so every reading is scaled by R/(R + 2r): 1995 parts per million at 0.35 Ω, 54029 at 10. Two more leads brought back from the bridge's own terminals, carrying only the instrument's input current, leave 3.5e-4 parts per million. Exciting with a current instead of a voltage does the same thing with no extra leads at all, because the lead resistance is in series with a source that does not care.

The two leads nobody counts

The leads that are in the bridge moved a lead out of the changing arm and turned a 999-microstrain error into nothing. The two leads carrying the excitation are still there, and they scale every reading: 0.35 ohms each on a 350-ohm bridge is 1,995 parts per million, exactly −2r/(R + 2r), the same on a full bridge as on a quarter one, and drifting 0.153 microstrain over twenty kelvin — twice what the three-wire fix left behind. Two more wires brought back from the bridge's own terminals leave 0.00035 parts per million. So does exciting the bridge with a current, which needs no extra wires at all.

applied · Bridge
A switching load's comb in a 10 mV reading: every tooth the same height. computed by solving, not by drawing. The harmonics of a 50-per-cent trapezoid of 100 mA at 100 kHz with a 10 ns edge, each multiplied by the shared conductor's impedance at its own frequency. Below 79.6 kHz the conductor is 5.00 mΩ of resistance and the teeth fall as 1/n. Above it the conductor rises as n while the harmonics still fall as 1/n, so the two cancel and every tooth from 239 kHz to 10.6 MHz is 0.828 mV — flat to 5.4 per cent across 52 of them. The comb stops at 31.8 MHz, which is the edge's own corner, so the total of 12.6 mV is decided by how many teeth there are: fitted over five edge rates it grows as the edge time to the power -0.497, against the −½ that a quadrature sum of equal teeth gives. A faster edge raises no harmonic and adds teeth. What a slow measurement keeps is a different term entirely: the zeroth harmonic, which is the 100 mA average through 5.00 mΩ of resistance — 500.0 µV, with no inductance in it and no edge, against the comb's 12.6 mV.

Every tooth the same height

A shared return conductor's impedance rises a decade per decade above 79.6 kilohertz and a switching load's harmonics fall a decade per decade, so the two cancel exactly: every harmonic of a hundred-milliamp square wave puts 0.828 millivolts into a ten-millivolt reading, from the conductor's own corner up to the edge's. The comb is flat rather than falling, so the total is decided by how many teeth there are — and since they add in quadrature it grows as the square root of the edge rate, 39.7 millivolts at a nanosecond against 4.03 at a hundred.

instruments · Common-impedance
Two sensors on one return: 511 µV in each reading, 100 nV in the difference. computed by solving, not by drawing. Two sensors referenced to the same local node, each reaching it through 2 mm of its own copper and each drawing 1 mA through that copper, with a 100 mA load returning through the 10 mm they all share. The interfering voltage is one voltage on one node, so it is added to both readings identically and the difference between them has none of it: 511.0 µV in each reading against 100.0 nV in the difference at a kilohertz, which is 74.2 dB. No resistor tolerance enters it — with the two stubs equal the rejection is exact to the solver, 166 decibels. What is left is the 10 per cent MISMATCH between the stubs times the sensors' own current, which the closed form puts at 100.0 nV, and it is proportional to the mismatch: 10.0 nV at 1%, 30.0 nV at 3%, 100 nV at 10%, 300 nV at 30%.

One voltage added to two readings

A difference amplifier across a shared return rejects somebody else's current by 54 decibels, and the number is four resistors rather than the amplifier. Two sensors referenced to the same node do better by a different mechanism entirely: the interfering voltage is one voltage on one node, so it is added to both readings identically and the difference between them has none of it — 511 microvolts in each reading against two and a half picovolts in the difference, which is 166 decibels and is exact rather than good. What is left is not a tolerance but a mismatch between the two sensors' own return paths, ten nanovolts per per cent of it.

instruments · Common-impedance
A star ground is a 0.50 mm daisy chain, not a point. computed by solving, not by drawing. 8 circuits of 100 mA each meeting at one pad, which reaches the plane through a via of 0.25 mΩ and 0.35 nH. The via is shared by every one of them, so the star's residual is 7 other currents through it — 175.0 µV at 1.00 kHz against the 1.800e+4 µV a daisy chain of 10 mm hops would give at its far end, a factor of 103. Measured as a length of the same track, the via's resistance is worth 0.500 mm and its inductance 0.350 mm — two different lengths, so a via has a corner of its own at 114 kHz where a track's is at 79.6 kHz wherever it is cut. The faint curves are other circuit counts. The two grow differently: fitted over counts from 2 to 32 the star's error goes as N^1.22 and the chain's as N^1.87, because a chain's hop carries every circuit beyond it — so the star's advantage grows with the count rather than shrinking. And a star has no position in it: a chain's far end reads 4.5 times its near end, which is a diagnostic no real fault should have.

A star is half a millimetre long

The repair for a shared return is a star ground, and it is described as making the shared length zero. It does not: every return meets at a pad, the pad reaches the plane through a via, and a via of a quarter-milliohm and 0.35 nanohenries is worth half a millimetre of one-ounce track by its resistance and 0.35 by its inductance — two different lengths, so a via has a corner of its own at 114 kilohertz where a track's is at 79.6. What that buys is a change of law rather than a factor: a daisy chain's error grows as the square of the circuit count and a star's as the first power, so the advantage runs from sixty times at two circuits to three hundred and forty at thirty-two.

instruments · Common-impedance
A cable's carrier limit is its conductors' matching, not its capacitance. computed by solving, not by drawing. A 350 Ω quarter bridge at 1000 µε, excited at a carrier frequency, with 10 metres of cable at 100 pF a metre on each output terminal. Each terminal sees the two arms meeting there in parallel against that capacitance, so each lags by arctan(2πf·(r/2)·C) with a corner at 909 kHz — and with the two conductors matched the lag CANCELS in the difference: -0.117° where each terminal has -5.71°. A mismatch does not cancel, and it arrives multiplied by the ratio of the excitation to the signal, which at 1000 µε is 2000: at 1 per cent the phase is 40.3° at a hundred kilohertz. So the excitation a 10-metre cable allows for 200 parts per million falls from 205 kHz matched to 1.82 kHz at 1 per cent. And it depends on the strain being measured — 183 Hz at 100 µε, 547 Hz at 300 µε, 1.82 kHz at 1000 µε, 5.45 kHz at 3000 µε — because the amplification is the reciprocal of the signal.

The carrier a cable allows

A carrier-frequency amplifier removes an instrument's offset drift by exciting the bridge with alternating current, and the cable that was a resistance at direct current becomes a capacitance. Both output terminals see the same arms and the same capacitance, so the lag cancels in the difference: ten metres matched leaves 0.15 degrees at a hundred kilohertz where each terminal has 6.3. What does not cancel is the mismatch between the two conductors, arriving multiplied by the ratio of the excitation to the signal — two thousand at a thousand microstrain — so one per cent of it takes the allowed carrier from 205 kilohertz to 1.8.

applied · Bridge
4 branches read with 10 Ω ammeters: one at a time every reading is low, all at once they add up and the small branches are high. computed by solving, not by drawing. A 1 A source into 4 parallel branches of 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, each read with an ammeter of 10 Ω. With one ammeter moved from branch to branch, each reading is low — by −0.989%, −0.902%, −0.099%, −0.010% — exactly r over the resistance that meter looks back into plus r, and the four readings add to 990.275 mA rather than 1000.000. With an ammeter in every branch at once the readings add to 1000.000 mA and are wrong by −0.892%, +7.939%, +8.910%, +9.008%: the meter in the largest branch pushes current out of it, and the small branches take it.

The readings that add up and are wrong

Four branches of 100 Ω, 1 kΩ, 10 kΩ and 100 kΩ share a 1 A source. Read them one at a time with a single 10 Ω ammeter moved from branch to branch, and every reading is within one per cent of the truth, but the four add to 990.275 mA. Put a 10 Ω ammeter in every branch at once and the four readings add to 1000.000 mA exactly, satisfying Kirchhoff's current law to the last digit. Yet three of them are 7.9 to 9.0 per cent high. The check a reader would run on the readings is passed by the wrong set and failed by the right one.

networks · Divider
A 1 ms pulse of 100 A warms the element 98.7 mK: 4.94 ppm at its end, 2.86 ppm averaged, against 775 ppm held. computed by solving, not by drawing, from the exact step response of the shunt's thermal network — element, terminations and board, 0.2, 4 and 15.8 K/W with 5 mJ/K, 0.1 J/K and 2 J/K — expanded into its Foster modes. A 100 A pulse of 1 ms through the 77.5 µΩ shunt the burden optimum picks, dissipating I·u* = 0.775 W: the element rises 98.7 mK by the pulse's end and cools after it. At 50 ppm/K that is 4.94 ppm of error at the end of the pulse and 2.86 ppm averaged over it, where the same current held long enough for the board to settle would read 775 ppm high.

The pulse that ends before the heat

A shunt's self-heating error is a fixed point in the steady state — 775 ppm at 100 A for the shunt the burden optimum picks, through 20 K/W. A pulse never reaches it. Heat leaves the element through a ladder network — the element's own 5 mJ/K, its terminations, the board — and a pulse short against the element's millisecond heats it adiabatically, by P·t over its heat capacity alone: 0.737 ppm for 100 µs at 100 A, a thousandth of the held figure, 10.7 ppm for 10 ms. Along the burden optimum the error is linear in the current, so the current a pulse may carry for 100 ppm is exact: 12.9 A held, 936 A for 10 ms, 13.6 kA for 100 µs. A reading averaged over the pulse carries half the error of one taken at its end while the heating is adiabatic, and the section-by-section thermal model a data sheet quotes is 10% wrong at 7 ms.

instruments · Current sensing
With the period 1.3% long, eight periods read a sine 0.599% wrong through a rectangle, 79.4 ppm through a triangle, 22.8 ppm through a Hann window. Integrated over each weighted window, the worst over every starting phase. The error in a sine's root-mean-square summed over N assumed periods 1.3% too long, against N, for four weights across the window. The rectangle is the plain sum: 0.640% at two periods and flat in N, δ/2 as the earlier measurement found. The triangle is flat in N too but 75 times smaller at eight periods. The Hann window falls with N, as the −2.67 power: 414 ppm at two periods, 0.293 ppm at thirty-two. The four-term Blackman–Harris is below 4.85 ppm at every count from two. A weight that falls to zero at the window's ends counts the extra sliver of period at almost nothing.

The window that forgives the period

A converter that sums a waveform's square over whole periods reads its RMS exactly, and one that sums over periods 1.3% too long reads a sine 0.6% wrong — δ/2, for any number of periods — because the extra sliver of period counts at full weight. Weight the window down to zero at its ends and the sliver barely counts. Over eight periods a triangular weight reads the same sine 79 ppm wrong and a Hann weight 23 ppm; and they are different kinds of cure. The triangle is the only one that changes the error's order, from first to second in δ, so it wins when the period is nearly right; the Hann weight keeps first order but falls with the number of periods, about as its inverse square, so it wins over long windows. Both cost nothing with the period exact — from two periods up. Over one, a Blackman–Harris weight reads even a sine 10% wrong. The price is noise: a Hann-weighted reading of noise scatters √1.5 as widely.

power · Rms and average
A tenth of a per cent of screen mismatch on the gauge's side cancels the pair's own phase: 18.2 kHz becomes 143 kHz. computed by solving, not by drawing. A 350 Ω quarter bridge at 1000 µε read through ten metres of screened pair: 100 pF a metre from each conductor to the screen, Cs, and 50 pF a metre between the two, Cp. The loss of the synchronously demodulated reading, 1 − cos φ, against the excitation frequency: for the capacitance between the conductors alone, for the screen capacitances alone with a mismatch of 0.1% on the gauge's side, for both together, and for a matched screen with nothing between. For a 200 ppm budget they allow 18.2 kHz, 18.3 kHz, 143 kHz and 205 kHz. The pair's capacitance lags the reading; extra screen capacitance on the gauge's own terminal leads it, by an amount divided by the signal, so at a mismatch of δ·Cp/Cs the two cancel to first order.

The capacitance that is already a difference

A screened pair has capacitance from each conductor to the screen, which cancels in a bridge's difference when the two match, and capacitance between the conductors, which is a difference from the start and cannot cancel. Ten metres of it between a 350 Ω quarter bridge and its carrier amplifier allows 18.2 kHz for a 200 ppm loss, against 205 kHz for the screen alone. The two phases are ωR·ΔC/δ and ωR·C, the same slope with one divided by the signal, so a screen mismatch of δ·Cp/Cs on the gauge's own terminal cancels the pair's lag exactly: 0.10% at 1000 µε takes the limit to 143 kHz. That is what a carrier amplifier's capacitance balance finds when it nulls the quadrature — and because the mismatch's phase changes sign with the strain and the pair's does not, the balance holds in tension above half the strain it was set at and never in compression.

applied · Bridge
Calibrated at the carrier, a matched cable allows 121 kHz where its phase allowed 16.5 kHz. computed by solving, not by drawing. A 350 Ω quarter bridge read through ten metres of screened pair, excited over leads of 0.35 Ω and 3 µH each with 500 pF between them. Solid: the worst error of a synchronous reading over strains from −2000 to +2000 µε after a calibration at +1000 µε at the operating carrier, with the reference taken from the source at the instrument; dashed, from the bridge's own terminals on sense wires; dotted, the phase measure 1 − cos φ the earlier readings of this cable used. For a 200 ppm budget: 16.5 kHz by phase, 119 kHz calibrated against the source and 121 kHz calibrated against the bridge.

What a calibration at the carrier removes

The phase a cable puts on a carrier-excited bridge is a loss of 1 − cos φ, and a loss that is the same at every strain is a gain error — which a shunt calibration at the operating carrier exists to remove. Calibrated there, a synchronous reading of a quarter bridge through ten metres of matched screened pair allows 121 kHz for 200 ppm over ±2000 µε, where the phase measure allowed 16.5; the capacitance between the conductors, which no matching could improve, stops mattering. Where the reference is taken — at the instrument or from sense wires at the bridge — moves that by 1.2 per cent. What calibration cannot touch is a screen mismatch: its leak is a fixed voltage, not a fixed gain, and a hundredth of a per cent of it takes 121 kHz back to 11.5.

applied · Bridge
3 relaxations over two decades come back within 0.22% from readings good to 1e-5. computed by solving, not by drawing. A one-per-cent tail divided equally among 3 branches spread over two decades from 0.1 ms, read at 80 times with 1e-5 of seeded noise on each reading, and the amplitudes recovered by least squares with the time constants given to the fit. Dots: one recovery; bars: the rms spread over 200. The condition number of the fit is 8.29e+0, and the worst branch comes back with an rms error of 0.224 per cent.

How many relaxations a settling curve holds

A settling curve read at eighty times, each good to ten parts in a million of the step, returns the amplitudes of three relaxations spread over two decades to 0.22 per cent and of eight to 46 — even when the fit is handed every time constant exactly. The condition number of the inversion grows by a factor of 7.7 for each extra relaxation packed into a decade, so each tenfold improvement in the reading buys a little more than one more relaxation a decade and no more. A settling measurement can count the dielectric's slowest relaxations; it cannot resolve a crowded spectrum, and how crowded is readable before any amplitude is.

transients · Doublet
A 1% mismatch in the cable reads as 3.0% of full-scale displacement, at every frequency. computed by solving, not by drawing. A capacitive displacement sensor — two 10 pF plates driven at ±5 V in antiphase facing a common electrode — read by a charge amplifier through two metres of three-conductor cable with 30 pF a metre between each excitation conductor and the sense conductor. With the two cross-capacitances mismatched by 1 per cent and the sensor centred, the reading's in-phase part (solid) is 3.00 per cent of full scale at every carrier from 100 Hz to 1 MHz: the cable is a sensor capacitance, and its mismatch a displacement. The quadrature part (dashed), which the cable's resistance makes, rises with frequency and a synchronous detector rejects it. With each conductor in its own screen the cross-capacitance is gone and so is the error.

The sensor that is the capacitance

Read a capacitive displacement sensor through the same kind of cable that limited a strain-gauge bridge's carrier, and the roles exchange. On the bridge, a 1% mismatch in the cable's capacitance was a quadrature leak that grew as the square of the carrier and could be escaped by a lower one. On a sensor of two 10 pF plates it is 3.0% of full-scale displacement, in phase, at every carrier from 100 Hz to 1 MHz — because a capacitance between an excitation conductor and the sense conductor is in parallel with the measurand and is a sensor capacitance. No detector phase separates it and no carrier escapes it; only geometry does. Screen each conductor separately and there is nothing left to mismatch.

applied · Bridge
A 10 mH winding read at 1 kHz while a 40 mH winding coupled to it at k = 0.99 is loaded from a short to an open. computed by solving, not by drawing at 61 loads for each coupling. Open, the first winding reads 10.0000 mH, its own inductance. Shorted, it reads 199.00 µH — L₁(1 − k²), the leakage seen from this side, because the second winding's current cancels all of the flux the two share. The change happens around a load equal to the second winding's reactance, 251 Ω. At k = 0.5, 0.9, 0.99 and 0.999 the shorted readings are 7500 µH, 1900 µH, 199.00 µH, 19.99 µH.

The leakage a series reading cannot see

The series-aiding and series-opposing readings recover a coupling coefficient to fifteen figures on a netlist, and on a bench with a 0.1% meter they recover it to about 0.001 at every coupling — which is useless exactly where transformers live. At k = 0.999 the leakage, L₁(1 − k²), is two thousandths of the winding, and the series route leaves it 133% uncertain: 88 of 400 seeded benches report a negative leakage. Shorting the other winding reads the leakage directly, to the meter's own 0.1%, and recovers k 943 times more precisely at 0.999. The short is a short only where the shorted winding's reactance dwarfs its resistance by 10k/√(1 − k²), and the winding to short is the one with the longer time constant.

magnetics · Mutual inductance
A reference 0° behind the excitation at 20 kHz, after a calibration at +1000 µε: matched, the worst error is 5.79 ppm; with a tenth of a per cent of screen mismatch, 0.607%. computed by solving, not by drawing: the error of a synchronous reading of a 350 Ω quarter bridge through ten metres of screened pair, against the strain read, with the reference 0° from the excitation at the bridge and a shunt calibration at +1000 µε through the same reference, for a matched screen (solid) and for 0.01 and 0.1 per cent of mismatch. The calibration takes out cos(φ − θ), because it multiplies the signal at every strain. It cannot take out a voltage that does not scale with the strain, and the mismatch's leak is such a voltage: a hyperbola through zero at the calibration strain. Worst over ±2000 µε: 0%: 5.79 ppm, 0.01%: 606 ppm, 0.1%: 0.607%.

The reference that should be late

A carrier amplifier's reference is taken from the excitation because that is what the output is meant to be in phase with, and nothing says that is where it belongs. A screen mismatch's leak arrives late by the common-mode node's lag and the difference's together, 525 ns on ten metres of pair and a 350 Ω bridge, so a reference on the excitation reads sin 3.7° of it at 20 kHz: 0.607 per cent of error at a tenth of a per cent of mismatch, and 72.6 parts per million with the reference 3.7° late. The null is a delay rather than a phase and does not move with the mismatch, and a delayed reference allows about five times the carrier on any mismatched screen. How exactly it must be held narrows as mismatch times carrier.

applied · Bridge
A shorted reading at k = 0.99: within 1% of the leakage from 279 Hz to 112 kHz, too high below through the winding's resistance and above through the leakage's resonance with 100 pF, at 1.13 MHz. computed by solving, not by drawing, in closed form, and by a netlist with a coupling element at three frequencies. The reading Im(Z)/ω at the driven terminals of a 10 mH and 40 mH pair, with the other winding shorted through its own 1 Ω, 0.5 Ω in the driven winding and 100 pF across it, against frequency, as a fractional distance from the leakage L₁(1 − k²) = 199.00 µH. Below the band the short does not cancel the shared flux. Above it the leakage and the capacitance resonate, at 1.13 MHz, and the reading rises to meet them. Band: 279 Hz to 112 kHz, a ratio of 402.1.

The window that moves

A shorted winding reads its own leakage only above the frequency at which the short really cancels the shared flux, and only below the leakage's resonance with whatever capacitance is across it. A tighter coupling raises the first edge as one over the root of 1 − k², and it was expected to pull the second down and close the window. It raises the second as well, by the same factor. So the band is 82 Hz to 36 kHz at k = 0.9, 279 Hz to 112 kHz at 0.99 and 2.81 kHz to 1.12 MHz at 0.9999, about four hundred times wide throughout. Its width is 0.01·L₂/(R₂k√(L₁C)), with the leakage nowhere in it, so what closes the window is the winding and not the coupling. Reading the real part as well is exact on paper and limited by phase.

magnetics · Mutual inductance

Named alongside it

The objects these essays reach for when they reach for this one.

ParasiticsDevice mismatchStrain gaugeWheatstone bridgeExcitationModel rangeCommon-mode rejectionClosed formCommon-impedanceLead resistanceReturn currentSynchronous detection