A conductor shared between two circuits, whose resistance and inductance put one circuit's current into the other's reading. It is the mechanism behind most of what looks like interference on a shared ground, and it is a millivolt problem rather than a volt one.

Named by 4 essays across 3 fields — each of them below, with the objects they name alongside it.

The millivolts in the wire, which are nobody's signal. computed by solving, not by drawing at 145 frequencies. A 10 mm run of one-ounce copper carries the return of a 100 mA load and the reference of a 10 mV sensor. Its 5.00 mΩ and 10.0 nH put 500.0 µV in series with the sensor at low frequency — 5.00% of the reading — rising a decade per decade above 79.6 kHz until at 1.59 MHz the error is the whole signal. The full solve and the interfering current times the shared impedance agree to 4.4e-7. The slider is the length of the shared run: both the resistance and the inductance are proportional to it, so every point on the curve moves down together and the corner stays at 79.6 kHz.

The millivolts in the wire

Ten millimetres of one-ounce copper is five milliohms and ten nanohenries, and if a hundred-milliamp load and a ten-millivolt sensor both return through it, half a millivolt of somebody else's current is added to the reading — five per cent of it, before anything has been amplified. Above 79.6 kilohertz the error rises a decade per decade with no ceiling, and shortening the shared run moves the whole curve down and the corner not at all.

instruments · Common-impedance
The same millivolts, subtracted — and what four resistors leave behind. computed by solving, not by drawing at 145 frequencies. A 10 mm run of one-ounce copper carries the return of a 100 mA load and the reference of a 10 mV sensor. Its 5.00 mΩ and 10.0 nH put 500.0 µV in series with the sensor at low frequency — 5.00% of the reading — rising a decade per decade above 79.6 kHz until at 1.59 MHz the error is the whole signal. The full solve and the interfering current times the shared impedance agree to 4.4e-7. The slider is the length of the shared run: both the resistance and the inductance are proportional to it, so every point on the curve moves down together and the corner stays at 79.6 kHz. The second curve is the same measurement made differentially: a unity-gain difference amplifier across the sensor, its four resistors carrying the worst-case skew 0.1% allows. It divides the error by 500.5 — 53.99 dB against a closed-form (1 + G)/(4·tolerance) of 53.98 dB — at every frequency, because four resistors have no frequency in them. What is left is 998.9 nV, which is not zero: the amplifier subtracts the interference and the resistors decide how much of it survives.

The rejection four resistors decide

Measuring a ten-millivolt sensor against a ground that somebody else's hundred milliamps is also using puts five hundred microvolts of their current into the reading. Subtracting the two ends of that conductor with a difference amplifier removes it — by 53.99 decibels with one-tenth-per-cent resistors, against a closed form of 53.98, which is a factor of five hundred and not a removal. The amplifier has nothing to do with it: the number is one plus the gain over four times the resistor tolerance.

instruments · Common-mode rejection
The bank reaches 1.500 mΩ and the load sees 141.5 mΩ at that same frequency. computed by solving, not by drawing. One bulk part and 20 ceramics, with a nanohenry of mounting loop each and two nanohenries of plane between the bank and the load, solved once per frequency and read at both nodes. The dashed curve is the bank's own node — what a probe on the parts measures. The solid one is the load. The bank's least impedance is 1.500 mΩ at 11.3 MHz, and at that frequency the load sees 141.5 mΩ, which is 94.3 times more, against 141.5 mΩ of plane reactance at that frequency. Whatever the parts do, the load's reading cannot fall below the reactance of the copper in front of them, and the parts reach their best by moving up the frequency axis into it.

The floor and the ceiling move apart

A decoupling bank is judged by two numbers — the lowest impedance it reaches and the highest frequency at which it still meets its target — and with no copper between the parts and the load both improve together as capacitors are added, 5.000 milliohms down to 1.500 and 19.8 megahertz up to 162. Three nanohenries of ordinary board separate them. The bank's own floor still falls 3.33 times while the load's falls 1.49, and the ceiling read at the parts climbs to 82.4 megahertz while the load's peaks at 8.06 and falls to 5.05. At twenty parts the two nodes disagree by a factor of 94 about the same solve.

power · Decoupling
One channel's load step reaches another through the supply, and the compensation decides by 4497×. computed by solving, not by drawing. Two identical amplifiers on one rail — sharing no signal node — with an ampere of load step pulled from the first and the second's output read. With the wiring left out the coupling is exactly zero, which is what the seven rungs below this one computed. With 30 nanohenries and fifty milliohms of rail and 10 microfarads of decoupling it is not: 3.84 microvolts per ampere at 271 kHz if the compensation capacitor returns to ground, and 17.29 millivolts per ampere at 2.33 MHz if it returns to the rail. That is a factor of 4497 decided by a modelling choice, which is why both are drawn. The channel that caused the step is unaffected: its own loop corrects the disturbance along with everything else, and the crosstalk is entirely a problem for the channel that did not.

The rail the load moves

Seven rungs of this ladder end by saying the same thing: the supply is an ideal voltage source, so a load step is drawn from a node that cannot be disturbed. Giving the rail an impedance turns out to change the disturbing channel's own output impedance by three parts in ten million — its loop corrects the supply along with everything else — and to open a path to a second amplifier that shares nothing with it but a wire. How large that path is is a modelling choice: 3.84 microvolts per ampere with the compensation capacitor returned to ground, 17.3 millivolts with it returned to the rail, a factor of four and a half thousand.

feedback · Capacitive load

Named alongside it

The objects these essays reach for when they reach for this one.

DecouplingLead resistanceLoadingModel rangeParasiticsReturn currentVerificationCapacitive loadCommon-mode rejectionComponent toleranceCrosstalkDifference amplifier