Finding a resistance small enough that the leads and contacts carrying the current are comparable with it. The burden voltage a shunt develops is the quantity to minimise, and the best one is a compromise between that and the noise it leaves.

Named by 3 essays across one field — each of them below, with the objects they name alongside it.

Measuring with 50 mΩ of lead in each wire. computed by solving, not by drawing at 61 resistances, twice each. The two-wire arrangement measures the leads too, so its error is 2×50 mΩ over whatever is being measured: one per cent at 10 Ω, and 10000% at 1 mΩ. The four-wire arrangement senses on a separate pair that carries almost no current, and its error stays under 1.0e-2% across the whole range.

Two terminals measure the leads as well

Fifty milliohms in each lead makes a two-wire measurement one per cent high at ten ohms, ten per cent high at one ohm, and a hundred per cent high at a tenth. Not approximately — the reading is the resistance plus the leads, and below about ten ohms most of what is being reported is the wire between the instrument and the thing.

instruments · Four-terminal
The best shunt drops 7.75 mV, whatever the current is. computed by solving, not by drawing. Two errors on one axis, both from solved networks: the shunt's own drop, which lowers the current that was to be measured, and the amplifier's 5.0 µV of offset divided by the voltage the shunt develops. The first rises with the burden voltage and the second falls, so the worst case has an interior minimum at 7.7460 mV — the geometric mean of the offset and the 12 V supply — where the error is 0.1291%, being twice the root of the offset over the supply. Neither the shunt's resistance nor the current appears in either number: at 1 A the answer is 7.75 mΩ, and at a hundred times the current it is the same burden voltage across a hundredth of the resistance. What does depend on the current is the 7.7 mW the shunt then dissipates, and 40 K of self-heating at 50 ppm/K is 0.2000% on its own. The third curve is the shunt's own Johnson noise in a kilohertz of measurement bandwidth, as a fraction of the current: it is 4.5e-6% at the best burden and is the only line here that moves with the current at all, falling as one over its square root — so above about an ampere it leaves the bottom of these axes entirely and is drawn nowhere rather than flattened onto the floor.

The ammeter that is a resistor

Every direct measurement of a current is a measurement of a voltage across something the current was made to flow through, so the instrument has two errors pointing opposite ways: a larger shunt changes the current, a smaller one leaves less for the amplifier's offset to be compared with. Written in the burden voltage they are the burden over the supply and the amplifier's offset over the burden, and the best of them is at the geometric mean — 7.75 millivolts on a twelve-volt rail, with a worst-case error of 0.129 per cent. Neither number contains a resistance, and neither contains the current: ten milliamps and a hundred amps want the same 7.75 millivolts.

instruments · Current sensing
Four wires against a 10 MΩ voltmeter. computed by solving, not by drawing at 81 resistances, twice each, with a voltmeter of 10 MΩ and 50 mΩ in every lead. The four-wire error is not zero: it is the voltmeter's own divider, −(R + 2R_lead)/(R + 2R_lead + R_m), which grows with the resistance being measured rather than shrinking. The two-wire error is that same quantity plus the leads, so it passes through zero at 1000 Ω — where the reading is right to 1.8e-12 while the four-wire reading is 0.0100% low — and above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.

The voltmeter four wires do not remove

A four-terminal measurement is described everywhere as removing the leads from the answer. It moves them. What is left is the voltmeter's own input resistance, and it grows with the resistance being measured rather than shrinking: with a ten-megohm voltmeter and fifty milliohms of lead, the four-wire reading is 0.0100 per cent low at a kilohm, where the two-wire reading is exactly right — 1.8 × 10⁻¹² — because its lead error and its loading error cancel. Above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.

instruments · Four-terminal