Low-resistance measurement — where it appears
Named by 4 essays across one field — each of them below, with the objects they name alongside it.
Two terminals measure the leads as well
Fifty milliohms in each lead makes a two-wire measurement one per cent high at ten ohms, ten per cent high at one ohm, and a hundred per cent high at a tenth. Not approximately — the reading is the resistance plus the leads, and below about ten ohms most of what is being reported is the wire between the instrument and the thing.
The ammeter that is a resistor
Every direct measurement of a current is a measurement of a voltage across something the current was made to flow through, so the instrument has two errors pointing opposite ways: a larger shunt changes the current, a smaller one leaves less for the amplifier's offset to be compared with. Written in the burden voltage they are the burden over the supply and the amplifier's offset over the burden, and the best of them is at the geometric mean — 7.75 millivolts on a twelve-volt rail, with a worst-case error of 0.129 per cent. Neither number contains a resistance, and neither contains the current: ten milliamps and a hundred amps want the same 7.75 millivolts.
The voltmeter four wires do not remove
A four-terminal measurement is described everywhere as removing the leads from the answer. It moves them. What is left is the voltmeter's own input resistance, and it grows with the resistance being measured rather than shrinking: with a ten-megohm voltmeter and fifty milliohms of lead, the four-wire reading is 0.0100 per cent low at a kilohm, where the two-wire reading is exactly right — 1.8 × 10⁻¹² — because its lead error and its loading error cancel. Above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.
The voltage four wires carry in
Four wires remove a resistance measurement's leads by keeping the sense loop free of current, and that loop is exactly where a thermoelectric voltage lives. Ten milliamps through a milliohm is ten microvolts, so every microvolt where a copper lead meets a brass terminal moves a single-polarity reading by a tenth. Reversing the current removes any EMF that holds still and leaves the change over one interval; the + − + delta leaves the second difference; four readings leave the third — 501 ppm, 4.14 and less at 10 mA while a disturbed junction settles. Against the shunt's own heating each scheme has a best current, and the least error rises only as the cube root of the shunt's temperature coefficient: the reading method moves the error by orders of magnitude, the alloy by a cube root.
Named alongside it
The objects these essays reach for when they reach for this one.
Four-terminal sensingDesign tradeoffKelvin connectionLoadingLead resistanceMeasurement conditionModel rangeTemperature coefficientGeometric meanInput impedanceInput offsetSelf-heating