Kelvin connection — where it appears
Named by 4 essays across 2 fields — each of them below, with the objects they name alongside it.
Two terminals measure the leads as well
Fifty milliohms in each lead makes a two-wire measurement one per cent high at ten ohms, ten per cent high at one ohm, and a hundred per cent high at a tenth. Not approximately — the reading is the resistance plus the leads, and below about ten ohms most of what is being reported is the wire between the instrument and the thing.
The leads that are in the bridge
A strain gauge on the end of two long wires cannot be told from a strain gauge under load: both leads in the changing arm is bit for bit the same netlist as a quarter bridge whose fractional change is larger by twice the lead over the gauge, which at 350 milliohms on a 350 ohm gauge is 999 microstrain that is not there. Moving one of those leads into the arm beside it leaves the output at exactly zero for every lead resistance drawn, and turns a twenty-kelvin drift of 76.8 microstrain into 0.077. That factor is two over the strain being read — 999 at a thousand microstrain and 9990 at two hundred — and it costs 0.100 per cent of sensitivity.
The voltmeter four wires do not remove
A four-terminal measurement is described everywhere as removing the leads from the answer. It moves them. What is left is the voltmeter's own input resistance, and it grows with the resistance being measured rather than shrinking: with a ten-megohm voltmeter and fifty milliohms of lead, the four-wire reading is 0.0100 per cent low at a kilohm, where the two-wire reading is exactly right — 1.8 × 10⁻¹² — because its lead error and its loading error cancel. Above 707 Ω the two-wire arrangement is the more accurate of the two at every resistance.
The two leads nobody counts
The leads that are in the bridge moved a lead out of the changing arm and turned a 999-microstrain error into nothing. The two leads carrying the excitation are still there, and they scale every reading: 0.35 ohms each on a 350-ohm bridge is 1,995 parts per million, exactly −2r/(R + 2r), the same on a full bridge as on a quarter one, and drifting 0.153 microstrain over twenty kelvin — twice what the three-wire fix left behind. Two more wires brought back from the bridge's own terminals leave 0.00035 parts per million. So does exciting the bridge with a current, which needs no extra wires at all.
Named alongside it
The objects these essays reach for when they reach for this one.
Four-terminal sensingLead resistanceExcitationLoadingLow resistance measurementMeasurement errorStrain gaugeTemperature coefficientWheatstone bridgeDesensitivityDesign tradeoffInput impedance